This treatment guidelines were developed to provide a consistently high level of safety, efficacy and comfort for your patients. It is important to perform a patient assessment prior to treatment,and consider factors such as the amount of laxity,volume, skin quality, age, and overall health of the individual. The treatment plan should also address the areas of greatest concern for the patient, and set realistic expectations for their anticipated outcome.

4.5mm Treatment Guideline

UPPER NECK
Treat the upper neck, making multiple passes and moving between columns as necessary, based on the patient’s response.
– Deliver the first line of energy in each column as close to the jaw bone (mandible) as possible, without hitting it. Align the transducer with the mandible, then slide it back and forth so the mandible “flashes” across the screen. Deliver energy when the mandible is not in the treatment window.
– Do not deliver energy below the line above the thyroid cartilage.
– Treating directly over the mandible or facial nerves may increase discomfort and the likelihood of potential side effects.
CHEEK
1. Treat the cheek region, making multiple passes and moving between columns as necessary, based on the patient’s response.
– Do not treat directly into the cheek bone or into the mandible.
– Ensure that the columns of treatment touch each other without overlapping, to ensure proper density.
– Treating directly over the mandible or facial nerves may increase discomfort and the likelihood of potential side effects.
LATERAL ORBIT
Align the transducer with the outside corner of the eye(closest to the ear),positioning the handpiece so the cord is towards the patient’s eye.
Deliver an equal number of lines above and below the marked line at the outside corner of the eye
– When treating, the end of the transducer closest to the ear will move more than the opposite side, creating a Fanning pattern.
– Reposition the transducer after delivering each line, instead of just pivoting on a point, to prevent retreating the same tissue.
BROW
1. Place the treat line of the transducer on the brow region, positioning the handpiece so the cord is towards the patient’s eye. Deliver the determined number of lines spread evenly in the region.
– Shorten the treatment line if the frontal bone is visible above the green line, if the shape of the frontal bone prevents good coupling or if the patient’s sensory response is significant.
GENERAL NOTES:
– Do not treat over the eye, in a location or with a technique,where the ultrasound energy may reach the eye.
– Do not deliver energy if bone is visible at or above the treatment line.
– Frequently verify proper coupling between the transducer and the skin.

NOTES: 3.0mm Treatment Guideline

UPPER NECK
1. Treat the upper neck, making multiple passes and moving between columns as necessary, based on the patient’s response.
– Deliver the first line of energy in each column as close to the jaw bone (mandible) as possible, without hitting it. Align the transducer with the mandible, then slide it back and forth so the mandible “flashes” acros the screen. Deliver energy when the mandible is not in the treatment window.
– Treating directly over the mandible or facial nerves may increase discomfort and the likelihood of potential side effects.
CHEEK
1. Treat the cheek region, making multiple passes and moving between columns as necessary, based on the patient’s response.
– Do not treat directly into the cheek bone or into the mandible.
– Ensure that the columns of treatment touch each other without overlapping, to ensure proper density.
– Treating directly over the mandible or facial nerves may increase discomfort and the likelihood of potential side effects.

LATERAL ORBIT
1. Align the transducer with the outside corner of the eye (closest to the ear),positioning the handpiece so the cord is towards the patient’s eye.
2. Deliver an equal number of lines above and below the marked line at the outside corner of the eye.
– When treating, the end of the transducer closest to the ear will move more than the opposite side, creating a fanning pattern.
– Reposition the transducer after delivering each line, instead

BROW
1. Place the treat line of the transducer on the brow region, positioning the handpiece so the cord is towards the patient’s eye. Deliver the determined number of lines spread evenly in the region.
– Shorten the treatment line if the frontal bone is visible above the green line, if the shape of the frontal bone prevents good coupling or if the patient’s sensory response is significant.

INFRAORBITAL
1 . Align the transducer on the mark below the eye , while gently pulling the infraorbital skin in a downward direction ( towards the feet ).
2. Deliver the first row of energy at the mark and then deliver each additional row below the last row delivered .Never tilt the transducer towards the orbit .

Treatment operation

Open system:
1.Determine the power cord on the back of the machine has been plugged in
2. Determine the rear of the machine main power switch is turned “ON“
3. Insert the key into the front of the instrument’s USB connector

Before treatment:
1. Add patient’s name and ID number in the system
2. Clean transducer
3. Click “Start Treatment” button

Start treatment:
1. Install the transducer on the handle
2. Applying gel on the transducer face then put transducer on the clean skin
3. Click “User Regions” button

Machine instruction and use instruction :
1. The software on the screen has the function of Identify transducer automatically . when you don’t install the transducer , the screen don’t have touch function . at the same time the hand holder S1 guiding light will twinkle . when you install the transducer , The software can identify transducer specification automatically and it will show on the relative area on the screen (including treatment depth and shot times and so on). S1 guiding light will twinkle slowly .
2 . After finish first step , under “standby” station , you can adjust the specification according to specific situation .Including “ energy (0.1 J‐2J) ” “ Length (10mm‐ 25mm)” . Then you can press“ User regions ” and enter into “ Ready ” station . If you want to alter the specification setting , please press “ End treatment ” and enter into “ standby ” station . then adjust the specification again .
3 . When the parameters are set , click the “User Regions” button , enter the Ready status ; For Easy to operate , you can just press B1 button on the handpiece , enter the Ready state , the blue light inside the round button will be on, tips ready . Then press the B2 button, the indicator will be on and cartridge will give out energy. Keep the cartridge on the face until the indicator light is off. That mean a shot is completed. DO NOT move the cartridge from face until the indicator light is off.

Operating Notice:
1. The operation person must be specially trained , should know well of the Human Face Structure ,and is proficient in the equipment operation .
2. Dividing the treatment area as well as avoiding the nerve sensitive areas.